The PR-730 uses 512 detectors to sample the emitted spectrum
from 380 to 780 nm yielding a resolution of 0.781nm/pixel.
Data is reported 1 nm. The spectral range of the PR-735 is 380
to 1100 nm for a resolution of 1.348 nm/pixel and reported in
2 nm increment. Up to 8 measurement apertures can be supplied
on the instrument providing unequaled spot size/sensitivity
flexibility. An external trigger circuit lets you make precise
measurements of flashing sources. The PR-730 can measure as
low as 0.0003 cd/m2 (0.00015 cd/m2 for the PR-735). Accurate
measurements of polarized sources like LCD's is quaranteed by
the extreamely low polarization characteristics (less than 0.2%).
Variable bandwidth capability - select between 2, 5, 8 nm spectral
bandwidths to insure the optimum conditions for virtually any
sample.
|